Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Integra Jarit GEMINI CLAMP 9 1/4 inch - used in Recalled by Integra LifeSciences Corp. Due to A small crack at the box lock corner...

Date: August 7, 2019
Company: Integra LifeSciences Corp.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Integra LifeSciences Corp. directly.

Affected Products

Integra Jarit GEMINI CLAMP 9 1/4 inch - used in vascular and cardiothoracic procedures used for clamping and occluding vessels. Cat. #140-332

Quantity: 237 units

Why Was This Recalled?

A small crack at the box lock corner may after multiple cleaning and sterilization cycles produce an increased risk of corrosion, weakening the instrument and resulting in premature need for replacement and/or breakage of the instrument

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Integra LifeSciences Corp.

Integra LifeSciences Corp. has 216 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report