Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Integra Camino Intracranial Pressure Monitoring Catheter with Licox IMC Bolt Recalled by Integra LifeSciences Corp. Due to Integra LifeSciences is recalling the Camino Intracranial Pressure...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Integra LifeSciences Corp. directly.
Affected Products
Integra Camino Intracranial Pressure Monitoring Catheter with Licox IMC Bolt Fitting, Catalog No. 110-4L The Camino Intracranial Pressure Monitoring Catheter is a sterile transducer-tipped pressure monitoring catheter with accessory items to be used as a diagnostic tool for rapidly determining and continuously monitoring intracranial pressure. The Camino catheter has a miniature transducer at the distal tip. This design eliminates the need for a fluid-filled system to carry pressure waves to an external transducer and thus, the problems inherent in such fluid-filled systems are eliminated. The Camino catheter is packaged with a zero adjustment tool which allows for the adjustment of the transducer during catheter preparation, and the LICOX Bolt Adapter Fitting. The Camino Intracranial Pressure Monitoring Catheter with Licox IMC Bolt fitting, Model 110-4L is indicated for use by a qualified Neurosurgeon when direct measurement of intracranial pressure in the parenchyma is clinically important. It is intended to be used only through a Licox Brain Oxygen Monitoring System Bolt.
Quantity: 1218 units
Why Was This Recalled?
Integra LifeSciences is recalling the Camino Intracranial Pressure Monitoring Catheters because they may have been incorrectly assembled. Specifically, the o-ring in the bolt adapter may be in the wrong location and as a result, there is a potential CSF/blood could leak from the bolt adapter.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Integra LifeSciences Corp.
Integra LifeSciences Corp. has 216 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report