Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Recalled by Integra LifeSciences Corp. Due to Integra became aware through a complaint from a...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Integra LifeSciences Corp. directly.
Affected Products
Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, 311 Enterprise Dr Plainsboro, NJ 08536 USA The Camino Advanced Monitor is a compact, portable device for use with Camino 110-4 series of Pressure / Temperature and Pressure Transducer-Tipped Catheters. Pressure and/or Temperature are measured at the Catheter tip, eliminating the need for external transducers, fluid, pressure tubing, and flush devices. The Camino displays intracranial pressure (ICP), intracranial temperature (ICT), and calculates cerebral perfusion pressure (CPP) (CPP = mean arterial pressure - mean intracranial pressure). The Camino provides a continuous display of the ICP waveform, as well as mean ICP, CPP, ICT, or systolic and diastolic waveform values. A continuous record of mean pressure and CPP values over the most recent 24-hour period is stored in memory, and can be displayed on command as a TREND either as the most recent 12 or 24 hour period. Outputs for ICP and ICT are available for use with patient bedside monitors. An isolated analogue output provides a continuous ICP waveform for hard copy documentation or data acquisition. An input to receive mean arterial pressure from a compatible patient bedside monitor is available for use when it is desired to have a CPP reading. The mean arterial pressure cannot be acquired from a non-invasive blood pressure cuff. Although the Camino is intended to be a stand-alone system, it also conveniently connects to hospital bedside monitoring systems.
Quantity: 275 CAM01 Monitors
Why Was This Recalled?
Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is used on CAM01 monitor that contains Mod 4 Firmware, the displayed values for the systolic and diastolic intracranial pressure (ICP) are reversed. But, if the CAM01 is used with an auxiliary bedside monitor, the bedside monitor displays these values correctly.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Integra LifeSciences Corp.
Integra LifeSciences Corp. has 216 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report