Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Cadence Size 5 Left Tibial Tray Implant Recalled by Integra LifeSciences Corp. Due to As a result of an internal review of...

Date: April 29, 2016
Company: Integra LifeSciences Corp.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Integra LifeSciences Corp. directly.

Affected Products

Cadence Size 5 Left Tibial Tray Implant, designed to treat ankle arthritis through replacement of the ankle joint with a prosthesis, thereby reducing pain, restoring alignment, and allowing for movement at the replaced joint. Catalog Number: 10207205; Lot Number #QJ0248

Quantity: 12 units

Why Was This Recalled?

As a result of an internal review of all labeling for the Cadence System, it was determined that the outer package label for the Size 5 Left Tibial Tray Implant reflects the material composition incorrectly. The labeling incorrectly reflects the material as Cobalt Chromium instead of Titanium.

Where Was This Sold?

This product was distributed to 2 states: NC, PA

Affected (2 states)Not affected

About Integra LifeSciences Corp.

Integra LifeSciences Corp. has 216 total recalls tracked by RecallDetector.

Related Recalls