Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

HemosIL ReadiPlasTin (20 mL) Recalled by Instrumentation Laboratory Due to The firm has received customer reports of performance...

Date: October 13, 2021
Company: Instrumentation Laboratory
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Instrumentation Laboratory directly.

Affected Products

HemosIL ReadiPlasTin (20 mL), Part number 0020301400. An in vitro diagnostic thromboplastin reagent.

Quantity: 5828 kits (310 US)

Why Was This Recalled?

The firm has received customer reports of performance issues with the affected lot, including increased imprecision, out of range quality controls, and prolonged sample results. If quality controls are not performed or do not pass for each vial of reagent, there is a potential risk of reporting an erroneous result.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Instrumentation Laboratory

Instrumentation Laboratory has 38 total recalls tracked by RecallDetector.

Related Recalls