Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Inspire Model 3028 Recalled by Inspire Medical Systems Inc. Due to There is a potential manufacturing defect which can...

Date: May 16, 2024
Company: Inspire Medical Systems Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Inspire Medical Systems Inc. directly.

Affected Products

Inspire Model 3028, IV Implantable Pulse Generator

Quantity: 24 devices

Why Was This Recalled?

There is a potential manufacturing defect which can lead to electrical leakage in the sensing circuit which causes a system malfunction and requires a revision surgery to replace the IPG,

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Inspire Medical Systems Inc.

Inspire Medical Systems Inc. has 4 total recalls tracked by RecallDetector.

Related Recalls