Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Roche Cobas 8000 and Cobas PRO Interface Module (CO8 IM)- Recalled by Inpeco S.A. Due to Firmware of the Interface Module with Roche Cobas...

Date: November 3, 2023
Company: Inpeco S.A.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Inpeco S.A. directly.

Affected Products

Roche Cobas 8000 and Cobas PRO Interface Module (CO8 IM)- IVD FlexLab Automation System -a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. Hardware versions: FLX-247-00 Firmware versions: COBASa_2.2.0, COBASa_2.3.0

Quantity: 1 unit US: 14 units OUS

Why Was This Recalled?

Firmware of the Interface Module with Roche Cobas 8000 and Cobas PRO (CO8 IM) may lead to a delayed sample tube processing.

Where Was This Sold?

This product was distributed nationwide across all 50 states.