Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Aptio Automation systems with Sysmex XN-9000/XN-9100 Interface Module (dual robot) Recalled by Inpeco S.A. Due to Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM)...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Inpeco S.A. directly.
Affected Products
Aptio Automation systems with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Clinical Chemistry Test Systems - Product Usage: The FlexLab 3.6 Automation is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The system consolidates analytical instruments, such as the ARCHITECT c8000 System into a unified workstation that perform a variety of instrument specific assays such as Sodium, Potassium and Chloride. Sodium, Potassium and Chloride measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance.
Quantity: 10 US and 22 OUS
Why Was This Recalled?
Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Inpeco S.A.
Inpeco S.A. has 33 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report