Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ImplantDirect Legacy(TM) 3 Implant Recalled by Implant Direct Sybron Manufacturing, LLC Due to One reported complaint and discovered that the outer...

Date: October 10, 2017
Company: Implant Direct Sybron Manufacturing, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Implant Direct Sybron Manufacturing, LLC directly.

Affected Products

ImplantDirect Legacy(TM) 3 Implant, 13MM L, REF 864713 This Legacy implant is a two-piece implant for one-stage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework.

Quantity: 75 units

Why Was This Recalled?

One reported complaint and discovered that the outer vial was out of round and undersized. Therefore a portion of the opening may not allow inner vial and components to slide freely out of the vial. As a result, the cap could possibly not be screwed on properly and the sterilization compromised.

Where Was This Sold?

This product was distributed nationwide across all 50 states.