Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MOSAIQ Oncology Information Systems interfaced to Elekta Treatment machines. An Recalled by Impac Medical Systems Inc Due to Impac Medical became aware of a problem with...

Date: June 8, 2012
Company: Impac Medical Systems Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Impac Medical Systems Inc directly.

Affected Products

MOSAIQ Oncology Information Systems interfaced to Elekta Treatment machines. An image enabled electronic medical record system used for oncology workflow management. Manufactured by: Impac Medical Systems 100 Mathilda Place, 5th floor Sunnyvale, Ca 94086

Quantity: 11 affected sites

Why Was This Recalled?

Impac Medical became aware of a problem with the recording of the Wedge Motor Units in MOSAIQ when it resumes after a machine terminates/faults, the remaining Wedge MU was being erroneously set to zero.

Where Was This Sold?

This product was distributed to 5 states: CT, GA, NJ, NY, VA

Affected (5 states)Not affected