Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Plum 360 Infusion System With MedNet/ Smart Card Plug And Play Module Recalled by ICU Medical Inc Due to A software malfunction of the infuser when used...

Date: August 1, 2018
Company: ICU Medical Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact ICU Medical Inc directly.

Affected Products

Plum 360 Infusion System With MedNet/ Smart Card Plug And Play Module, Model # 30010. Infusion pump.

Quantity: 47,152 devices

Why Was This Recalled?

A software malfunction of the infuser when used with a MedNet Custom Drug Library (CDL). When a new CDL is downloaded and installed on the infuser the infuser can become inoperable and show a blank screen on the User Interface.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About ICU Medical Inc

ICU Medical Inc has 50 total recalls tracked by RecallDetector.

Related Recalls