Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Plum 360 Infusion System Recalled by ICU Medical Inc Due to Potential for the Connectivity Engine Module of the...

Date: June 15, 2017
Company: ICU Medical Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact ICU Medical Inc directly.

Affected Products

Plum 360 Infusion System

Quantity: 68,559 units

Why Was This Recalled?

Potential for the Connectivity Engine Module of the affected unit to disengage from the main chassis, causing the unit to shut down without an alarm notification. This would delay initiation of therapy or interrupt an infusion.