Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ICU Medical Plum Solos Recalled by ICU Medical, Inc. Due to Internal testing found that Plum Duo pumps with...

Date: April 22, 2025
Company: ICU Medical, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact ICU Medical, Inc. directly.

Affected Products

ICU Medical Plum Solos, Item number 400011001; infusion pump

Quantity: 214 units

Why Was This Recalled?

Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About ICU Medical, Inc.

ICU Medical, Inc. has 592 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report