Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Triathlon Prim CEM FXD BPLT 7 Recalled by Howmedica Osteonics Corp. Due to The internal profile of the Triathlon Prim (Baseplate)...

Date: July 1, 2021
Company: Howmedica Osteonics Corp.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Howmedica Osteonics Corp. directly.

Affected Products

Triathlon Prim CEM FXD BPLT 7, Catalog No. 5520B700

Quantity: 14 US; 2 OUS

Why Was This Recalled?

The internal profile of the Triathlon Prim (Baseplate) is potentially oversized; there may be a witness mark (taper) at the bottom of the internal profile of the baseplate, which prevents the insert from locking on the baseplate.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Howmedica Osteonics Corp.

Howmedica Osteonics Corp. has 229 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report