Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

AccuLIF TL Insertion Handle The AccuLIF Insertion Handle attaches to Recalled by Howmedica Osteonics Corp. Due to When the lever of the Insertion Handle is...

Date: September 8, 2017
Company: Howmedica Osteonics Corp.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Howmedica Osteonics Corp. directly.

Affected Products

AccuLIF TL Insertion Handle The AccuLIF Insertion Handle attaches to the implant and is specifically designed to allow a surgeon to insert the implant into the disc space with a fixed engagement between the two components to permit a controlled positioning of the implant in the anterior portion of the intervertebral space, and then through activation of the rotational pivot between the Handle and implant for final placement of the crescent-shaped cage across the anterior portion of the disc space. The AccuLIF TL Cages are indicated for intervertebral body fusion with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DOD) at one level or two contiguous levels from L2 to S1.

Quantity: 130

Why Was This Recalled?

When the lever of the Insertion Handle is in the "locked" position, the implant may be able to pivot rather than maintaining a rigid connection.

Where Was This Sold?

Nationally

About Howmedica Osteonics Corp.

Howmedica Osteonics Corp. has 229 total recalls tracked by RecallDetector.