Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

The Plum A+ is a dual-line volumetric infusion system designed Recalled by Hospira Inc. Due to Firm's investigation determined that the Proximal Occlusion Alarms...

Date: March 25, 2016
Company: Hospira Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Hospira Inc. directly.

Affected Products

The Plum A+ is a dual-line volumetric infusion system designed to meet the growing demand for hospital-wide, as well as alternate site and home healthcare, standardization. With its primary line, secondary line, and piggyback fluid delivery capability, the Plum A+ is suited for a wide range of medical/surgical and critical care applications. Full compatibility with LifeCare Plum Series administration sets and accessories. Each system includes a pumping module (hereafter called the infuser) and an assortment of disposable IV sets (hereafter called a set), and optional accessories. The Plum A+ host device contains a Connectivity Engine peripheral module that provides wired Ethernet and wireless 802.11 a/b/g local area networking capabilities. This allows the Hospira Mednet networked application software to download drug libraries to the infuser and enable the autoprogramming feature.

Quantity: 2567 Plum A+ infusion pumps.

Why Was This Recalled?

Firm's investigation determined that the Proximal Occlusion Alarms observed in the field were attributed to the mechanism calibration process being performed with an out of specification metal cassette, resulting in an offset in proximal pressure.

Where Was This Sold?

U.S. distribution : Natrionwide. Foreign :Brazil, Canada, Chile, Colombia, Ecuador, Guatemala, Mexico, Oman, Peru, Taiwan, Turkey, United Kingdom and Uruguay.

About Hospira Inc.

Hospira Inc. has 245 total recalls tracked by RecallDetector.

Related Recalls