Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

The GemStar Pump is a small and lightweight Recalled by Hospira Inc. Due to During infusions of 2.0 mL/hr the motor assembly...

Date: March 21, 2013
Company: Hospira Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Hospira Inc. directly.

Affected Products

The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA alkaline batteries. When powered by batteries, The GemStar Pump is ideal for ambulatory patients.

Quantity: 1) List Number 13086: 286 pumps; 3) List Number 13087: 45,376 pumps; 4) List Number 13088: 23,492 pumps

Why Was This Recalled?

During infusions of 2.0 mL/hr the motor assembly may rotate backwards capturing additional medication resulting in over-infusion.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Hospira Inc.

Hospira Inc. has 245 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report