Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

LifeShield LATEX-FREE HemoSet 100 mL Burette I.V. PlumSet Recalled by Hospira Inc. Due to Hospira identified an incorrect lower lid (set component)...

Date: February 25, 2014
Company: Hospira Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Hospira Inc. directly.

Affected Products

LifeShield LATEX-FREE HemoSet 100 mL Burette I.V. PlumSet, Nonvented, 105 inch with Prepierced Reseals, OPTION-LOK and Inline Dual Channel Cassette. For administration of blood and blood bags for use with Plum Series Infusers.

Quantity: 3,893 sets

Why Was This Recalled?

Hospira identified an incorrect lower lid (set component) had been supplied and was being used during the manufacturing process of list number 11241-03, a HemoSet containing a burette and blood filter, intended for infusion of blood/blood products. In a gravity delivery, the correct lower lid dispenses 15 drops per mL and the incorrect lower lid found dispenses 10 drops per mL.

Where Was This Sold?

This product was distributed to 14 states: FL, IN, IA, KS, KY, ME, MO, MT, PA, TN, TX, UT, VA, WA

Affected (14 states)Not affected