ABX PENTRA Creatinine 120 CP Recalled by Horiba Instruments, Inc dba Horiba Medical Due to This recall was initiated because some customers were...

Date: September 14, 2012
Company: Horiba Instruments, Inc dba Horiba Medical
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Horiba Instruments, Inc dba Horiba Medical directly.

Affected Products

ABX PENTRA Creatinine 120 CP, Model Number Creatinine Part # A11A01868, Lot Numbers: 203273706, 203273707, 203273808, 203273709. The ABX PENTRA Creatinine 120 CP is an in vitro diagnostic assay for the quantitative determination of creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffe method). It is composed of a 27 ml monoreagent cassette. Reagent is a chemical solution with additives. ABX PENTRA Creatinine 120 CP reagent, with associated calibrator and controls, is a diagnostic reagent for quantitative in vitro determination of Creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffe method). Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. This reagent is intended for use on the ABX PENTRA 400 Clinical Chemistry Analyzer.

Quantity: 8,503 units

Why Was This Recalled?

This recall was initiated because some customers were experiencing an alarm during Creatinine calibration "REAG RANGE HIGH". This alarm caused the calibration to fail and prevented the user from using the Creatinine assay.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Horiba Instruments, Inc dba Horiba Medical

Horiba Instruments, Inc dba Horiba Medical has 17 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report