Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

10 ML REAGENT CUP Recalled by Horiba Instruments, Inc dba Horiba Medical Due to The ABX PENTRA Reagent Containers do not always...

Date: November 21, 2017
Company: Horiba Instruments, Inc dba Horiba Medical
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Horiba Instruments, Inc dba Horiba Medical directly.

Affected Products

10 ML REAGENT CUP, B1034626, SAP 1221034626 for use with ABX PENTRA 400 / PENTRA C400

Quantity: 10 bags, 100 units each

Why Was This Recalled?

The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.

Where Was This Sold?

This product was distributed nationwide across all 50 states.