Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Hitachi Echelon MRI System/Oasis MRI Recalled by Hitachi Medical Systems America Inc Due to Hitachi discovered a software error with a feature...

Date: June 15, 2012
Company: Hitachi Medical Systems America Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Hitachi Medical Systems America Inc directly.

Affected Products

Hitachi Echelon MRI System/Oasis MRI, Software. These are software controlled MRI devices. Product Usage: The Echelon and Oasis MR systems are imaging devices, intended to provide the physician with physiological and clinical information that can be useful in diagnosis determination.

Quantity: 165 units

Why Was This Recalled?

Hitachi discovered a software error with a feature called Multi-Planar Reconstruction (MPR). MPR takes an original volume dataset reconstructed in any plane view and creates a new image set from another view (i.e. axial to coronal, coronal to sagittal, etc.). The error causes the first image in the dataset to have a right/left reversed orientation. The software error can occur randomly when execu

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Hitachi Medical Systems America Inc

Hitachi Medical Systems America Inc has 17 total recalls tracked by RecallDetector.