Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MHI-TM2000 Linear Accelerator System Product Usage: MHI-TM2000 Linear Accelerator System Recalled by Hitachi Ltd., Medical System Operations Group Due to Due to a system controller software anomaly, the...

Date: December 8, 2017
Company: Hitachi Ltd., Medical System Operations Group
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Hitachi Ltd., Medical System Operations Group directly.

Affected Products

MHI-TM2000 Linear Accelerator System Product Usage: MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

Quantity: 1

Why Was This Recalled?

Due to a system controller software anomaly, the patient positioning deviation correction may not be applied and may result in the wrong part of the patient being irradiated.

Where Was This Sold?

This product was distributed to 1 state: NY