Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

These are communication systems Recalled by Hill-Rom Inc. Due to An unrecoverable error may occur where a component...

Date: October 14, 2021
Company: Hill-Rom Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Hill-Rom Inc. directly.

Affected Products

These are communication systems, labeled as: a. NaviCare Nurse Call, Part Numbers: 3.6, 3.7, 3.9 b. Voalte Nurse Call, Part Number 4.0 *Product was rebranded from NaviCare to Voalte in 2020

Quantity: 1,050 units

Why Was This Recalled?

An unrecoverable error may occur where a component will fail to complete the reboot cycle, become non-functional and must be replaced.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Hill-Rom Inc.

Hill-Rom Inc. has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report