Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Liko Overhead Rail System Recalled by Hill-Rom, Inc. Due to Liko/Hill-Rom to date has received two reports from...

Date: October 31, 2012
Company: Hill-Rom, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Hill-Rom, Inc. directly.

Affected Products

Liko Overhead Rail System, Ceiling Mounted Rail Systems for all Liko Stationary Lifts including Likorall, Multirall and Masterlift Systems Product Usage - The Overhead Rail System allows overhead lifts to be mounted to the ceiling allowing versatility for use in lifting situations. It can be used with different Liko ceiling lifts.

Quantity: 3816 in total

Why Was This Recalled?

Liko/Hill-Rom to date has received two reports from facilities alleging that the overhead rail system has failed to the extent that the overhead rails have fallen down while in use in high humidity (swimming pool) areas. Analysis of the incidents suggest that the components (hardware) used to secure the rail to the ceiling had corroded and may be unsuitable for environments with an atmosphere con

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Hill-Rom, Inc.

Hill-Rom, Inc. has 35 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report