Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Medtronic HVAD Pump Implant Kits Recalled by Heartware, Inc. Due to Pumps manufactured with impellers from a subset of...

Date: November 19, 2020
Company: Heartware, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Heartware, Inc. directly.

Affected Products

Medtronic HVAD Pump Implant Kits, labeled as: a. PUMP 1103, b. PUMP 1104, c. PUMP 1104JP, d. PUMP 1100, e. Model 1101, f. Model 1102, g. Model 1104CA-CLIN, h. Model 1205, i. Model MCS1705PU

Quantity: 19,562 units

Why Was This Recalled?

Pumps manufactured with impellers from a subset of lots from a single supplier are failing to initially start, restart, or experiencing a delay in restart within the maximum number of restart attempts at a rate substantially higher than pump in the overall population.

Where Was This Sold?

This product was distributed nationwide across all 50 states.