Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

i.v. Station and i.v. Station ONCO Automated Drug Compounding System Recalled by HEALTH ROBOTICS S.R.L. Due to Mislabeling

Date: November 23, 2016
Company: HEALTH ROBOTICS S.R.L.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact HEALTH ROBOTICS S.R.L. directly.

Affected Products

i.v. Station and i.v. Station ONCO Automated Drug Compounding System, Pharmacy Compounding System Device Listing Number E454448

Quantity: 121 systems

Why Was This Recalled?

The firm has initiated a product correction due to a bag carousel motor error causing misalignment and resulting in a mislabeling of final IV bag preparations.

Where Was This Sold?

Shipped throughout United States.

About HEALTH ROBOTICS S.R.L.

HEALTH ROBOTICS S.R.L. has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report