Sensei X Robotic Catheter System Recalled by Hansen Medical Inc Due to The Set Up Joint subassembly of the Sensei...

Date: December 4, 2014
Company: Hansen Medical Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Hansen Medical Inc directly.

Affected Products

Sensei X Robotic Catheter System, model numbers 02057 and 08257. - Set up joint (SUJ)mechanical clamp that secures the SUJ to the catheter laboratory table. Product Usage: The Sensei X Robotic Catheter System is an electromechanical software controlled system designed to facilitate navigation and control of Hansen Medicals robotically steerable catheters for collecting electrophysiological data within the heart atria with electro-anatomic mapping and recording systems.

Quantity: 59 USA, 40 OUS

Why Was This Recalled?

The Set Up Joint subassembly of the Sensei X Robotic Catheter System may become loose relative to the operating table with the potential for the device to fall causing, uncontrolled catheter movement. cardiac tampenade, delayed procedure, or crush injury to the patient and/or user.