Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ReTurn 7500/7500 Aid Recalled by Handicare USA, Inc. Due to During the period of April 2009  May...

Date: April 15, 2014
Company: Handicare USA, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Handicare USA, Inc. directly.

Affected Products

ReTurn 7500/7500 Aid, transfer

Quantity: 29 units , plus 47 wing handles were shipped as spare parts

Why Was This Recalled?

During the period of April 2009  May 2011, Handicare AB used a wing handle in which the screw and handle, after repeated tightening, may slide apart. This causes a small amount of play in the attachment of the ladder, which in turn results in high load at the attachment holes of the ladder. Eventually, this may cause mechanical fatigue in the material of the ladder.

Where Was This Sold?

This product was distributed to 8 states: CT, IL, ME, NE, NV, NJ, NY, PA

Affected (8 states)Not affected

About Handicare USA, Inc.

Handicare USA, Inc. has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report