Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Hamilton-T1 Recalled by Hamilton Medical, Inc. Due to Degrading capacitors on the control board of ventilators...

Date: July 26, 2023
Company: Hamilton Medical, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Hamilton Medical, Inc. directly.

Affected Products

Hamilton-T1, REF: 161009, Intensive Care and Transport Ventilator

Quantity: 1216 units

Why Was This Recalled?

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Hamilton Medical, Inc.

Hamilton Medical, Inc. has 19 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report