Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Ventilator HAMILTON-C6 Recalled by Hamilton Medical AG Due to Ventilator may enter sensor fail mode, ventilation may...

Date: May 15, 2024
Company: Hamilton Medical AG
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Hamilton Medical AG directly.

Affected Products

Ventilator HAMILTON-C6, PN: 160021

Quantity: 68

Why Was This Recalled?

Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube for open suctioning, 3)Sensor error initiated, sensor fail mode occurs, 4)Patient is re-connected with sensor fail mode active, 5) Ventilation is not re-initiated by the ventilator.

Where Was This Sold?

This product was distributed nationwide across all 50 states.