Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

HAMILTON-C6 Intensive Care Ventilator Recalled by Hamilton Medical AG Due to Ventilator status indicator board can become loose, which...

Date: June 27, 2022
Company: Hamilton Medical AG
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Hamilton Medical AG directly.

Affected Products

HAMILTON-C6 Intensive Care Ventilator, REF: 160021

Quantity: 358

Why Was This Recalled?

Ventilator status indicator board can become loose, which could lead to water ingress (disinfectants) that may lead to technical fault alarms. Multiple technical faults in a short time may force ventilators into Safety Ventilation (blower runs constantly) or Ambient Sate (inspiratory channel/expiratory valves opened; patient breaths room air unassisted) with Panel connection lost message displayed

Where Was This Sold?

This product was distributed nationwide across all 50 states.