Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Hamilton-C3 Ventilator Unit Intended to provide ventilator support to adults Recalled by Hamilton Medical AG Due to After two and a half years, the battery...

Date: March 12, 2018
Company: Hamilton Medical AG
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Hamilton Medical AG directly.

Affected Products

Hamilton-C3 Ventilator Unit Intended to provide ventilator support to adults and pediatrics and optionally infants and neonates in the hospital.

Quantity: 739 units total

Why Was This Recalled?

After two and a half years, the battery fuel gauge may indicate a higher battery charge than what is available.

Where Was This Sold?

This product was distributed to 29 states: AZ, CA, CO, FL, GA, HI, IL, IN, KY, MA, MI, MN, MO, MT, NE, NV, NC, ND, OH, OK, PA, SC, SD, TN, TX, UT, VT, VA, WA

Affected (29 states)Not affected

About Hamilton Medical AG

Hamilton Medical AG has 12 total recalls tracked by RecallDetector.

Related Recalls