Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Hamilton-C1 Ventilator Unit Product Usage: The HAMILTON-C1 ventilator is intended Recalled by Hamilton Medical AG Due to An out of specification component on the control...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Hamilton Medical AG directly.
Affected Products
Hamilton-C1 Ventilator Unit Product Usage: The HAMILTON-C1 ventilator is intended to provide positive pressure ventilation support to adults and pediatrics. The device is used in the in the intensive care ward, in the recovery room, and during transfer of ventilated patients within the hospital.
Quantity: 16 devices
Why Was This Recalled?
An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking alarm LED on the device s front side.
Where Was This Sold?
This product was distributed to 9 states: CA, CO, IL, LA, MN, NY, OK, VA, WI
About Hamilton Medical AG
Hamilton Medical AG has 12 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report