Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Haag Streit Surgical Floor stands FS 2-11 (Ref no. 615H511) Recalled by Haag-Streit USA Inc Due to Software error -Software REF 615 588 versions 2.0...

Date: November 20, 2020
Company: Haag-Streit USA Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Haag-Streit USA Inc directly.

Affected Products

Haag Streit Surgical Floor stands FS 2-11 (Ref no. 615H511) and FS 2-15 (Ref no. 615H515) in combination with the operating microscopes HS Hi-R NEO 900 (Ref no. 657 820) and HS Hi-R NEO 900A (Ref no. 657 821), with software REF 615 588 versions 2.0 to 3.3 - Product Usage: use in ophthalmology and in other disciplines with vertical direction of view

Quantity: 3 units US

Why Was This Recalled?

Software error -Software REF 615 588 versions 2.0 to 3.3, movement of the focusing is triggered by pressing the button of the footswitch it might occur that the movement does not stop when releasing the button, result in potential risk to the patient's eye

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Haag-Streit USA Inc

Haag-Streit USA Inc has 9 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report