Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Torque Limiting Adapter: Torque Limiting Attachment Recalled by Greatbatch Medical Due to Greatbatch Medical has initiated a global Field Recall...

Date: April 26, 2016
Company: Greatbatch Medical
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Greatbatch Medical directly.

Affected Products

Torque Limiting Adapter: Torque Limiting Attachment, Non Sterile, Distributed by Zimmer. PRECIMED, Torque Limiting Adaptor, Non Sterile.. Customer Number/Model: 00-2360-080-00 T10232, T116245 T116245, TLA025-25 TLA025-25, 00-2360-080-05 T10233, MAC21003 T116430, TLA025-30 TLA025-30, MAC21001 T112972, MAC21001 T13621, TLA025-20 TLA025-20.

Quantity: 5952

Why Was This Recalled?

Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device.

Where Was This Sold?

This product was distributed to 3 states: IN, MI, TX

Affected (3 states)Not affected

About Greatbatch Medical

Greatbatch Medical has 16 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report