Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Gold Cup Leadwire and Snap Leadwire with the following component Recalled by GN Otometrics Due to The electrode lead connector has a protruding conductor...

Date: August 1, 2019
Company: GN Otometrics
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GN Otometrics directly.

Affected Products

Gold Cup Leadwire and Snap Leadwire with the following component descriptions: 1. 24" Snap Leadwire, Blue; 2. 40" Snap Leadwire, Blue; 3. 24" Snap Leadwire, Red; 4. 40" Snap Leadwire, Red; 5. 24" Snap Leadwire, Green; 6. 40" Snap Leadwire, Green; 7. 24" Snap Leadwire, Brown; 8. 40" Snap Leadwire, Brown; 9. 24" Snap Leadwire, Orange; 10. 40" Snap Leadwire, Orange; 11. 24" Snap Leadwire, White; 12. 40" Snap Leadwire, White; 13. 24" Snap Leadwire, Black; 14. 40" Snap Leadwire, Black; 15. 24" Snap Leadwire, Yellow; 16. 40" Snap Leadwire, Yellow; 17. 24" Snap Leadwire, Set of 5; 18. 24" Snap Leadwire, Set of 7; 19. 40" Snap Leadwire, Set of 5; 20. 40" Snap Leadwire, Set of 7; 21. 24" Snap Leadwire, Set of 5; 22. 40" Snap Leadwire, Set of 5; 23. 48" Gold Cup Leadwires, Blue; 24. 48" Gold Cup Leadwires, Red; 25. 48" Gold Cup Leadwires, Green; 26. 48" Gold Cup Leadwires, Brown; 27. 48" Gold Cup Leadwires, Orange; 28. 48" Gold Cup Leadwires, White; 29. 48" Gold Cup Leadwires, Black; 30. 48" Gold Cup Leadwires, Yellow; 31. 48" Gold Cup Leadwires, Set of 5; 32. 48" Gold Cup Leadwires, Set of 8; Product Usage: The snap electrode leads are used in conjunction with disposable electrodes and the Gold Cup electrodes are reusable electrodes and leads. Both types of electrodes and leads are used for evoked potential testing for audiological, neurological and/or vestibular testing. The electrode leads are connected to the main device by the electrode lead connector and the electrode is attached to the patient contact end of electrode lead.

Quantity: 4,882 units (41950 individual electrode leads supplied In 5,7 & 10 packs)

Why Was This Recalled?

The electrode lead connector has a protruding conductor beyond insulation of approximately 0.1 mm.

Where Was This Sold?

This product was distributed nationwide across all 50 states.