Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Behind-the-Ear hearing aids Recalled by GN Hearing A/S Due to GN Hearing received reports from our manufacturing site...

Date: September 18, 2019
Company: GN Hearing A/S
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GN Hearing A/S directly.

Affected Products

Behind-the-Ear hearing aids, Model Numbers: Q962-DRW, T561-DRW, T562-DRW, T762-DRW, T777-DW, AMZ964-DRW, ET588-DWH, ET598-DW, ET788-DWH, ET798-DW, ET988-DWH, ET998-DW, FT861-DRW, FT862-DRW, RE561-DRWC, RE762-DRW, RE961-DRWC, RE962-DRW, PZ862-DRW, LT561-DRW, LT561-DRWZ, LT562-DRW, LT588-DW, LT761-DRW, LT761-DRWZ, LT762-DRW, LT777-DW, LT788-DW, LT961-DRW, LT961-DRWZ, LT962-DRW, LT967-DW, LT977-DW, LT988-DW, TST1763-DRW, TST1764-DRW, TST663-DRW, TST963-DRW, TST964-DRW, VA861-DRW, VA861-DRWZ, VA862-DRW

Quantity: 590 units

Why Was This Recalled?

GN Hearing received reports from our manufacturing site in US regarding a software bug in our service & repair tool which affected the capture and restoration of user data in a limited number of repaired hearing aids. The software bug caused these repaired hearing aids to contain altered specifications for sound amplification, potentially putting users residual hearing at risk of experiencing unintended additional available gain to varying degrees i.a.18dB flat linear gain or potentially in worst case Full On Gain. However, no injuries have been reported.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About GN Hearing A/S

GN Hearing A/S has 2 total recalls tracked by RecallDetector.