Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Getinge 9100-Series Washer Disinfector models: 9120E Recalled by Getinge Usa Sales Inc Due to Inadequate documentation verifying whether device installation has been...

Date: April 1, 2022
Company: Getinge Usa Sales Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Getinge Usa Sales Inc directly.

Affected Products

Getinge 9100-Series Washer Disinfector models: 9120E, 9125E, 9127E, 9128E, 9122EW, 9128EW.

Quantity: 127

Why Was This Recalled?

Inadequate documentation verifying whether device installation has been executed in accordance with the manufacturer's requirements. This can lead to installation failure modes and hazardous situations.

Where Was This Sold?

This product was distributed to 33 states: AZ, CA, CO, CT, FL, GA, ID, IL, IN, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NJ, NY, NC, ND, OH, OK, PA, SC, SD, TX, VA, WA, WV, DC

Affected (33 states)Not affected

About Getinge Usa Sales Inc

Getinge Usa Sales Inc has 48 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report