Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Flow-e Anesthesia System Recalled by Getinge Usa Sales Inc Due to Potential cybersecurity vulnerability that could lead to a...

Date: December 11, 2023
Company: Getinge Usa Sales Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Getinge Usa Sales Inc directly.

Affected Products

Flow-e Anesthesia System, model 6887900

Quantity: 86 (US)

Why Was This Recalled?

Potential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous situation.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Getinge Usa Sales Inc

Getinge Usa Sales Inc has 48 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report