Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Recalled by Getinge Usa Sales Inc Due to The power backup battery, under certain conditions, may...

Date: November 7, 2022
Company: Getinge Usa Sales Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Getinge Usa Sales Inc directly.

Affected Products

Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)

Quantity: 65 US

Why Was This Recalled?

The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power failure. This could require the clinician to use manual emergency ventilation of another ventilator is not available and the backup battery is exhausted.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Getinge Usa Sales Inc

Getinge Usa Sales Inc has 48 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report