Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Washer Disinfector Aquadis 56 Recalled by Getinge Disinfection Ab Due to The frequency inverter is not programmed by supplier...

Date: April 2, 2025
Company: Getinge Disinfection Ab
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Getinge Disinfection Ab directly.

Affected Products

Washer Disinfector Aquadis 56; Model Number: 56A

Quantity: 102 units (8 US, 94 OUS)

Why Was This Recalled?

The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Getinge Disinfection Ab

Getinge Disinfection Ab has 7 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report