Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used Recalled by GE OEC Medical Systems, Inc Due to mA Accuracy may exceed the design and labeling...

Date: September 15, 2014
Company: GE OEC Medical Systems, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE OEC Medical Systems, Inc directly.

Affected Products

GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.

Quantity: 61 units installed in US

Why Was This Recalled?

mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).