Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

GE HC OEC 9900 Elite and GE OEC 9800 Fluoroscopes Recalled by GE OEC Medical Systems, Inc Due to GE OEC recalled certain 9800 Fluoroscope & 9900...

Date: April 26, 2010
Company: GE OEC Medical Systems, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE OEC Medical Systems, Inc directly.

Affected Products

GE HC OEC 9900 Elite and GE OEC 9800 Fluoroscopes Snubber Boards are a PCB with a 70 amp fuse to provide fused protection of internal system components from the high voltage system components, such as the X-ray tube or High Voltage Transformer. System model number A349855, Snubber Board number 00-880405-03 rev 5. This product is used for image intensified fluoroscopic x-rays for interventional radiologic procedures.

Quantity: 845 units

Why Was This Recalled?

GE OEC recalled certain 9800 Fluoroscope & 9900 Elite Models because the Snubber Board did not conform to the IEC 60601-1 standard for clearance and creepage of voltage isolation.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About GE OEC Medical Systems, Inc

GE OEC Medical Systems, Inc has 48 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report