Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

GE Nuclear Medicine 600/800 series systems Recalled by GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING Due to A mitigation may not be correctly implemented. If...

Date: December 19, 2022
Company: GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING directly.

Affected Products

GE Nuclear Medicine 600/800 series systems, specifically: a) 850, b) 850 / Dismount, c) 850 3/8 inch NM Gantry, d) 860 / Dismount, e) 860 3/8 inch NM Gantry, f) 860 5/8 inch NM Gantry, g) 860 BJ, h) NM 830 3/8 inch BJ, i) NM/CT 850 3/8 inch BJ, j) NM/CT 850 3/8 inch ROW, k) NM/CT 850 3/8 inch ROW_, l) NM/CT 850 CT 850 3/8 inch, m) NM/CT 850 ES 3/8 inch, n) NM/CT 850 ES 3/8 inch ROW, o) NM/CT 860 3/8 inch, p) NM/CT 860 3/8 inch BJ, q) NM/CT 860 3/8 inch ROW, r) NM/CT 860 ADVANCE, s) NM/CT 860 advance 3/8 inch NM Gantry, t) NM/CT 860 BEYOND, u) NM/CT 860 beyond 3/8 inch NM Gantry, v) NM/CT 860 China for China, w) NM/CT 860 CT 860 3/8 inch, x) NM/CT 860 ES 3/8 inch, y) NM/CT 860 ES 3/8 inch ROW, z) NM/CT 850 3/8 inch, aa) NM/CT 850 ES 3/8 inch ROW, bb) NM/CT 860 3/8 inch CH, cc) XELERIS 4 DR WORKSTATION; Emission Computed Tomography System

Quantity: 380 units

Why Was This Recalled?

A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING