Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

GE Healthcare Revolution Apex Expert Recalled by GE Medical Systems, LLC Due to GE HealthCare has become aware of a potential...

Date: February 18, 2025
Company: GE Medical Systems, LLC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Medical Systems, LLC directly.

Affected Products

GE Healthcare Revolution Apex Expert, System, X-ray, Tomography, Computed

Quantity: 42 units

Why Was This Recalled?

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

Where Was This Sold?

Worldwide distribution.

About GE Medical Systems, LLC

GE Medical Systems, LLC has 146 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report