Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

GE Healthcare Carestation 620 A1 Recalled by GE Medical Systems China Co., Ltd. Due to GE HealthCare has become aware that certain Carestation...

Date: March 21, 2025
Company: GE Medical Systems China Co., Ltd.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Medical Systems China Co., Ltd. directly.

Affected Products

GE Healthcare Carestation 620 A1, Model/REF Number 1012-9620-200

Quantity: 383 units

Why Was This Recalled?

GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.

Where Was This Sold?

This product was distributed nationwide across all 50 states.