Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Discovery MR750w Nuclear Magnetic Resonance Imaging System labeled as the Recalled by GE Medical Systems, LLC Due to GE Medical Systems has discovered a small area...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact GE Medical Systems, LLC directly.
Affected Products
Discovery MR750w Nuclear Magnetic Resonance Imaging System labeled as the following product name sold under: Product Name (Model Number): A. 3.0T MR750 32CH (911000-3); B. 3.0T MR750W (5911000-3, 5481490-2, 5499460-7, 5499460-10, 5373011-3, 5352293, 5911000-3); C. 3.0T MR750W GEM (5499460-10, 5481490-2); D. 3.0T MR750W GEM ENAB (5911000-3); E. 3.0T MR750W GEM ENABLED (5352293, 5481490-2, 5481490-2, 5396469-3, 2394952-100, 5911000-3, 5373011-3); F. 3.0T MR750W GEM ENABLED PLU (5352293); G. 3.0T MR750W GEM ENABLED PLUS (5352293, 5499460-10); H. 3.0T MR750W W GEM (5352293, 5911000-3); I. 3.0T MR750W WITH GEM (5481490-2, 5911000-3); J. 3-0T MR750 32 CHANNEL (5352293); K. 3-0T MR750W FUTURE GEM CAPABLE (5352293, 5911000-3, 5499460-7); L. 3-0T MR750W GEM (5495558, 5352293); M. 3-0T MR750W GEM ENABLED (5911000-3, 5352293, 2394952-100, 5481490-2, 5499460-7, 5373011-3, 5495558, 5481490, 5396469-3, 5481490-3); N. 3-0T MR750W GEM ENABLED PLUS (5911000-3, 5352293, 5499460-10, 5481490-2); O. 3-0T MR750W NON GEM (5911000-3); P. 3T 750W (5921000-2, 5911000-3, 5396469-3, 5352293, 5481490-2, 5481490-3); Q. DISCOVERY MR 750W 3.0T (5481490-2, 5911000-3, 5499460-10); R. DISCOVERY MR750W (5911000-3); S. DISCOVERY MR750W 3 (5481490-2, 5911000-3); T. DISCOVERY MR750W 3T (5911000-3, 5396469-3, 5373011-3, 5481490-2); U. DISCOVERY MR750W GEM (5396469-3, 5352293); V. DISCOVERY MR750W WITH GEM (5911000-3); W. MR 3.0T MR750 W (5352293); X. MR 3.0T MR750W GEM ENABLED (5911000-3, 5352293); Y. MR 3-0T MR750W (5911000-3); Z. MR DISCOVERY 750W GEM (5911000-3); AA. MR OPTIMA MR750W 3.0T (5911000-3); BB. MR750W (5911000-3); CC. MR750W GEM ENABLED (5911000-3, 5499460-7); DD. MR750W GEM ENABLED PLUS (5352293); EE. MR750W PGR 32-CHANNEL (5911000-3); FF. 3.0T MR750W GEM ENAB (5911000-3) Product Usage: The Discovery MR750w 3.0T system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the Discovery MR750w 3.0T system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance..
Quantity: 516 Units Total
Why Was This Recalled?
GE Medical Systems has discovered a small area on the bore surface of the Discovery MR750w Nuclear Magnetic Resonance Imaging System, normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About GE Medical Systems, LLC
GE Medical Systems, LLC has 146 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report