Anesthesia device service kits. ASSY-MSN Recalled by GE Medical Systems, LLC Due to The Drive Gas Check Valve of certain Anesthesia...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact GE Medical Systems, LLC directly.
Affected Products
Anesthesia device service kits. ASSY-MSN, VENT ENGINE ASSY, 1503-8101-000 ASSY-MSN, VENT ENGINE ASSY 7900, 1009-8216-000 ASSY-MSN, VENT ENGINE ASSY AESPIRE 7900 1009-8423-000 ASSY-MSN, VENT ENGINE ASSY, AESTIVA 7900 MRI 1503-8102-000 ASSY-MSN, VLV ASSY DR GAS CHK BCG, Manufacturing assembly - Buy 1503-3006-000 Service kits containing the ventilator Drive Gas Check Valve shipped from April 20, 2015 through October 2015 are also affected and could have been installed on Aespire 7900, Aespire View, Aestiva 7900, Aestiva MRI, Aisys, Aisys CS2, Avance, Avance CS2 Anesthesia, and Amingo devices. (Service kit part numbers: 1009-8216-000, 1503-3006-000, 1503-8102-000, 1009-8423-000, 1503-8101-000).
Quantity: 240
Why Was This Recalled?
The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About GE Medical Systems, LLC
GE Medical Systems, LLC has 146 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report