Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

v|tome|x L Series non-destructive testing cabinet x-ray systems Recalled by Ge Inspection Technologies Llc Due to It was discovered during testing of the v/tome/x...

Date: September 29, 2020
Company: Ge Inspection Technologies Llc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ge Inspection Technologies Llc directly.

Affected Products

v|tome|x L Series non-destructive testing cabinet x-ray systems

Quantity: N/A

Why Was This Recalled?

It was discovered during testing of the v/tome/x L450 system s/n PA1790, that the pre-warning signals does not provide "Audible and visible warning signals within the cabinet" if the "warm up" mode was not completed after initializing the system when first turned on.