Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

GE Inspection Technologies x|tome|x L240 an Cabinet X-ray system Recalled by GE Inspection Technologies, LP Due to It was discovered that the upper left corner-shaped...

Date: January 13, 2015
Company: GE Inspection Technologies, LP
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact GE Inspection Technologies, LP directly.

Affected Products

GE Inspection Technologies x|tome|x L240 an Cabinet X-ray system

Quantity: 2

Why Was This Recalled?

It was discovered that the upper left corner-shaped lead shield in the window in one of the doors of the v|tome|x L240 installed had lost adhesion and fall out of position, which could lead to gaps in shielding.

Where Was This Sold?

This product was distributed to 2 states: OH, PA

Affected (2 states)Not affected